The post Edible Oils Adulterant Analysis appeared first on NIR-For-Food.
]]>
The following samples were acquired for the study: reference EVOO and refined olive oil from Sigma-Aldrich, several authentic EVOO samples from California, commercially available EVOO from Italy, various refined vegetable oils from local grocery stores, and two palm olein samples from Indonesia and Thailand. NIR spectra were collected using different FT-NIR spectrometers from the same vendor using a diffuse reflectance probe with a liquid attachment and a 2 mm pathlength. All spectra were collected at 8 cm-1 resolution and six replicate spectra were collected for each sample and averaged into one spectrum. Before calibration work, tests were conducted to assess the effects of external changes on the aforementioned absorption bands. The same EVOO sample was measured for over an hour at room temperature and it was noted that the absorption band at 5280 cm-1 decreased while the other band was unchanged. To confirm this change was because of temperature, an EVOO sample was heated at 50°C for ten minutes and spectra were acquired over the ten minute span. Next, two tests were run to see if the decrease in absorption was due to the loss in volatile compounds. A vacuum at 260 mBar was applied to an EVOO sample for 50 minutes and rescanned, showing a significant decrease in the 5280 cm-1 band. Likewise, bubbling nitrogen gas was passed through an EVOO sample for fifty minutes and after rescanning, a similar result was observed. A decrease in the 5280 cm-1 band was also observed after scanning spiked samples of EVOO with fully refined olive oil, corn oil, vegetable oil, and palm olein. The band at 5180 cm-1 showed little or no change with heat, vacuuming, bubbling with nitrogen gas, or the addition of other oils. The last test added 10 µL of water to an EVOO sample to ensure that changes in the two bands were not occurring because of changes in moisture. After scanning the sample, no spectral change was observed in the two bands of interest after the water added.
After the external tests were conducted, chemometric software was used to choose wavenumber ranges that included the characteristic features of the two bands of interest. In order to eliminate the need to measure and integrate the area of the two absorption bands, a Partial Least Squares (PLS) calibration model was created. For all samples, the normalized ratio of the two integrated band areas was calculated. The sample with the highest value (about 1.7 to 1) was arbitrarily assigned a value of 100 for FT-NIR index. California EVOO samples ranged from 100 to 92 on the scale. The certified EVOO sample from Sigma Aldrich was 90 while the refined olive oil was 40. Commercial EVOO from Italy ranged from 86 to 97. To expand the scale and more accurately estimate the lower end, a synthetic triolein was scanned and a very low FT-NIR index score of 5 was determined.
The second portion of this study used prepared gravimetric mixtures of authentic EVOO spiked with nine common adulterants. The adulterants used are: soybean oil, sunflower oil, corn oil, canola oil, hazelnut oil, high oleic acid safflower oil, peanut oil, palm olein, and refined olive oil. The first eight listed here were mixed in concentrations from 3% to 30% by weight adulterant and in the case of refined olive oil, up to 60% total weight was used. Initial analysis indicated that one universal PLS model for all adulterants was not feasible. However, the fatty acid profiles for all adulterants was examined and it was determined that spitting the adulterants into four groups based on fatty acid profiles might show better results. This was the case after creating four PLS models from the four groups. Shown below are the results for the FT-NIR index and four adulterant PLS models.
| FT-NIR Index | R² = 0.995 | RMSEP= 1.7 |
| R² = 0.999 | RMSEP= 0.9% w/w |
| R² = 0.995 | RMSEP= 2.2% w/w |
| R² = 0.999 | RMSEP= 1.0% w/w |
| R² = 0.976 | RMSEP= 3.7% w/w |
| Ligurian Samples | 92.5% |
| Non-Ligurian Samples | 81.5% |
Rapid Characterization of Transgenic and Non-Transgenic Soybean Oils by Chemometric Methods Using NIR Spectroscopy – Luna, da Silva, Pinho, et al., Spectrochimica Acta Part A 100 (2013) 115-119
Genetic engineering of food has become advanced in recent years but in many parts of the world such food is considered undesirable, creating the need for rapid screening methods for proper labeling. Forty transgenic and forty non-transgenic soybean oil retail samples were procured for the study. Transmission spectra were collected from 1100 nm to 2500 nm at 1 nm intervals. Five spectra were collected per sample and averaged into one spectrum. Various post-processing algorithms were applied to the spectra and classification methods were performed to determine the best method for separating the two types of samples. The best results were obtained using Support Vector Machine-Discriminant Analysis (SVM-DA), a technique used for binary classification. This technique corrected predicted 100% of the non-transgenic samples and 90% of the transgenic samples in the validation set. Since there were no reference tests conducted on the samples and the classification was based strictly on what the retail label displayed, it is possible that some of the samples were misclassified to start. However, the results shown in this study are good enough for screening purposes to classify transgenic and non-transgenic soybean oils.
https://www.ncbi.nlm.nih.gov/pubmed/16131099
Classification and Quantification of Palm Oil Adulteration Via Portable NIR Spectroscopy – Basri, Hussain, Bakar, et al., Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy 173 (2017) 335-342
Lard can produce cheap oil and can be mixed with other vegetable oils as an adulterant. A NIR spectrometer was set up to operate in both transmission and transflectance mode for comparative purposes. Samples were prepared of pure palm oil, pure lard, and mixtures at various increments of added lard from 0.5% to 50%. Absorbance spectra of all samples were collected in both modes from 950 nm to 1650 nm and exported for chemometric analysis. Classification analysis was performed on pure vs. adulterated palm oil and Partial Least Squares (PLS) regression models for quantifying lard in palm oil was created from both sets of NIR spectra.
| Classification Accuracy | 0.95 |
| Lard PLS | R² = 0.9998 |
| Classification Accuracy | 0.93 |
| Lard PLS | R² = 0.9996 |
The post Edible Oils Adulterant Analysis appeared first on NIR-For-Food.
]]>The post Fish Oil Analysis appeared first on NIR-For-Food.
]]>Fish oils are known for their benefits to human health because they contain Omega-3 fatty acids (n-3 FA), which are families of long-chain polyunsaturated fatty acids (PUFAs). Many such fatty acids have essential roles in the prevention and treatment of coronary heart disease, diabetes, autoimmune disorders, arthritis, hypertension, and inflammatory diseases. The two predominant fatty acids with these benefits are Eicosapentaenoic Acid (EPA, 20:5 n-3) and Docosahexaenoic Acid (DHA, 22:6 n-3) and these are found almost exclusively in seafood products. Dietary guidelines recommend 0.5 g to 1.0 g of these fatty acids per day and more for various treatments. Fish oil contains about twenty percent of their total fatty acids as PUFA, making them highly susceptible to both oxidation and hydrolytic degradation of lipids. These reactions can cause rancid odors and can even give rise to the formation of products like aldehydes, hydroperoxides, and epoxides that can detract from the nutritional content. Some fish oils are converted to biodiesel by alkali or lipase-catalyzed transesterification reactions and Free Fatty Acids (FFA) content is a critical parameter in these reactions. Moisture, FFA, Peroxide Value (PV), Anisidine Value (AV), EPA, and DHA are monitored using physical and chemical techniques, but these methods are slow and expensive to implement, especially in an industrial process setting. There is a pressing need to determine these parameters that cannot only test large amounts of samples but do so in a cheap and timely fashion. One such method that has been examined is NIR spectroscopy.
Measurement of chemical parameters for quality control purposes has been studied using NIR spectroscopy for fish oils. The results of most studies have been promising. One such study examined monitoring the oxidative and hydrolytic degradation of lipids in fish oil. Good correlation was shown between NIR spectral data and models for FFA and moisture. The correlation was worse for PV and AV, but this may have occurred due to an error in the reference method. Measuring these parameters with accuracy using NIR spectroscopy has been proven in studies using other types of edible oils. Another study compared using NIR, IR, and Raman spectroscopy to measure EPA, DHA, and Total n-3 FA. Another study used a variety of fish oil supplements and IR spectroscopy to measure EPA and DHA with excellent results. FTIR was examined as an alternative for replacing a conventional titration method to measure FFA in fish oils intended for biodiesel production and the results proved the feasibility of this measurement. However, implementing NIR spectroscopy as a large-scale testing method for fish oil will require extensive calibration work to include different varieties and incorporation of natural sources of variability because they are natural products.
Multivariate Determination of Free Fatty Acids and Moisture in Fish Oils by Partial Least-Squares Regression and Near-Infrared Spectroscopy – Cozzolino, Murray, Chree, Scaife – Science Direct LWT 38 (2005) 821-828
One hundred and sixty fish oil samples from a fishmeal factory were scanned in transflectance mode using a NIR monochromator instrument. Wavelength range was from 1100 nm to 2500 nm. The fish oil samples had different chemical compositions due to differences in species and seasonality as well as storage in different tanks. Reference values for Free Fatty Acids, Moisture, Peroxide Value, and Anisidine Value were obtained during the production of material and not at the time of scanning. There was a minimum of five days between sampling and scanning time.
| Free Fatty Acids (FFA) | R2= 0.96 |
| Moisture | R2= 0.94 |
| Peroxide Value (PV) | R2= 0.60 |
| Anisidine Value (AV) | R2= 0.93 |
Good correlation was obtained for FFA and moisture calibration models and validation predictions proved the feasibility of the models for measuring these parameters. However, the PV and AV models did not work as well for the validation set. In the case of PV, the delay in time between collecting reference data and scanning the samples contributed to the low correlation coefficients as some oxidation almost certainly occurred in the samples, creating high reference error. In the case of AV, the correlation coefficient was high, but many of the samples had a reference value of zero as secondary oxidation had not yet occurred. Some validation predictions showed a value less than zero which invalidates the model in a real-time setting. PV and AV have been accurately predicted in other types of edible oil and should be able to in fish oil as well with a more carefully constructed calibration set. The results do prove the feasibility of monitoring hydrolytic degradation of lipids in fish oil using NIR spectroscopy.
https://www.sciencedirect.com/science/article/pii/S0023643804002865
Determination of Omega-3 Fatty Acids in Fish Oil Supplements Using Vibrational Spectroscopy and Chemometric Methods – Bekhit, Grung, Mjos, Applied Spectroscopy, Volume 68, Number 10, 2014
A Fourier Transform Infrared (FT-IR), Near-Infrared (NIR), and Raman spectrometer were all used to scan sixty-one fish oil supplements to predict concentrations of Eicosapentaenoic Acid (EPA), Docosahexaenoic Acid (DHA), and Total Omega-3 Fatty Acids (n-3 FAs). GC was used as a reference method to determine concentrations of these three parameters. EPA and DHA concentrations were expressed as percentages relative to the total mass of fatty acids. Total n-3 FAs were the sum of concentrations of eight single n-3 FAs.
| Scanning Parameters | |
|---|---|
| FT-IR | 4000 cm-1– 650 cm-1, 32 scans per average, 4 cm-1 resolution, ATR Crystal background |
| NIR | 1100 nm -2500 nm, 2 nm intervals, 32 scans per average, 4 cm-1 resolution, adjustable transflectance probe, 1 mm pathlength |
| Raman | 3450 cm-1 – 0 cm-1, 5 second exposure time, <500mW excitation laser at 785 nm |
| Results | ||
|---|---|---|
| FT-IR | EPA- | R2= 0.994, Range = 1820 cm-1 – 650 cm-1 |
| DHA- | R2= 0.983, Range = 1820 cm-1– 650 cm-1 | |
| Total n-3 FAs- | R2= 0.985, Range = 3090 cm-1 -2800 cm-1 & 1790 cm-1 –650 cm-1 | |
| NIR | EPA- | R2= 0.979, Range = 1530 nm -1900 nm |
| DHA- | R2= 0.972, Range = 1530 nm -1940 nm | |
| Total n-3 FAs- | R2= 0.997, Range = 1630 nm -1870 nm | |
| Raman | EPA- | R2= 0.977, Range = 1800 cm-1-769 cm-1 |
| DHA- | R2= 0.966, Range = 1800 cm-1-769 cm-1 | |
| Total n-3 FAs- | R2= 0.993, Range = 3450 cm-1 – 0 cm-1 | |
Models were created for all three instruments using different pre-processing techniques and selective wavelength ranges. The best results are shown above. Excellent correlation was shown for all parameters, demonstrating the potential to measure EPA, DHA, and Total n-3 FAs in fish oil supplement. These three vibrational spectroscopy methods have distinct advantages and disadvantages depending on the situation. While FT-IR is easy to implement in a laboratory, the small pathlength when reflecting off an ATR crystal makes it ill-suited for some measurements. Likewise, Raman is unable to detect Raman shifts in certain molecules. NIR is best suited to a process environment because no sample preparation is needed.
Application of Infrared Spectroscopy for Characterization of Dietary Omega-3 Oil Supplements – Plans, Wenstrup, Rodriguez-Saona, J Am Oil Chem Soc (2015) 92:957-966
Commercial Omega-3 dietary supplements of fish oil, cod liver oil, and flaxseed oil from different manufacturers were procured for this study. Duplicate samples for some of the supplements were also purchased at a later date, ensuring separate lots from the original supplements. All samples were scanned from 4000 cm-1 to 700 cm-1using 4 cm-1resolution and an ATR crystal as a background. GC was performed to create a Fatty Acid profile for each sample. Calibration models were created for seven fatty acids using selective wavenumber ranges.
| C14:0 | R2= 0.99 |
| C16:0 | R2= 0.99 |
| C16:1 | R2= 0.95 |
| C18:1 | R2= 0.95 |
| EPA (Eicosapentaenoic Acid) | R2= 0.99 |
| DHA (Docosahexaenoic Acid) | R2= 0.99 |
| FFA (Free Fatty Acids) | R2= 0.96 |
Results showed a consistent classification of four groups of samples. Based on EPA/DHA content, oil source, and factors associated with processing (FA alkyl ester or triglyceride). The samples in the grouping showed clear spectral distinction in the absorption bands for triglycerides and FA alkyl esters. Models showed excellent performance, especially for EPA and DHA which are the two primary Omega-3 fatty acids of interest in such supplements. Prediction results confirm the capability to estimate the main FAs of fish oil supplements.
https://rd.springer.com
FTIR Determination of Free Fatty Acids in Fish Oils Intended for Biodiesel Production – Aryee, Van de Voort, Simpson, Process Biochemistry 44 (2009) 401-405
Biodiesel is commonly derived from vegetable oils and animal fats by alkali or lipase-catalyzed transesterification reactions. Free Fatty Acids (FFA) is a critical parameter in the conversion of fish oil to methyl esters and the feasibility of using FTIR for determining FFA was examined. The performance of FTIR was assessed as an alternative to the conventional AOCS titration method. Spectra were collected using a transmission flow cell at a resolution of 8 cm-1. Sixteen scans were averaged per spectrum and ratioed against an open-beam background spectrum.
The FTIR method involves simultaneously extracting FFAs, conversion to salts using a weak base, and measuring the carboxylate band (COO-) at 1573 cm-1. The method was found to respond linearly to oleic acid addition, producing a calibration equation for FFA. Validation samples proved the FTIR method was more accurate and reproducible than the titrimetric method.
https://www.sciencedirect.com/science
The post Fish Oil Analysis appeared first on NIR-For-Food.
]]>The post Cooking Oil Analysis appeared first on NIR-For-Food.
]]>Frying with cooking oil is a widespread technique for preparing foods. The frying process degrades oil because it creates a series of chemical reactions, such as hydrolysis and thermo-oxidative degradation, that reduce nutritional contents and form undesirable compounds. Thus, it is essential from both an economic and health perspective for producers and consumers to use oil that is capable of withstanding repeated frying cycles. Many parameters can be used to measure the quality of cooking oil, and the most frequently measured include acidity, peroxides, total polar materials (TPM), and oxidative stability index (OSI). Used cooking oil also has uses in manufacturing, such as being the carbon source in the production of polyhydroxyalkanoates (PHA). This occurs through fermentation in a bioreactor. The current reference methods for both cooking oil quality and reaction monitoring are expensive, time-consuming, and labor-intensive, creating the need for a fast, cheaper alternative for monitoring the parameters of interest. One such method that has been examined is NIR spectroscopy.
Measurement of chemical parameters for quality control purposes has been studied using NIR spectroscopy for cooking oil. One study examined measuring parameters of interest for thermo-oxidative degradation of cooking oil. Numerous reactions take place while cooking oil is used for frying and they are grouped into oxidative, hydrolytic, and thermal degradation. The correlation between calibration models of some parameters used to monitor these reactions and the reference method was good enough to show NIR spectroscopy as a potential replacement for traditional reference tests. Another study examined monitoring a bioreactor producing a PHA using used cooking oil as a carbon source. NIR spectroscopy was used to monitor the fermentation and results were good enough to show it is a suitable method for online monitoring and assistance of bioreactor control.
Evolution of Frying Oil Quality Using Fourier Transform Near-Infrared (FT-NIR) Spectroscopy – Calero, Munoz, Perez-Marin, et al., Applied Spectroscopy 2018, Vol. 72(7) 1001-1013
Fourteen types of vegetable oil were used and were subjected to successive frying processes. After each frying, a sample was scanned for NIR spectra, and reference tests were performed for the parameters of interest. Spectra were collected using a transflectance probe from 12500 cm-1 to 4000 cm-1 using 8 cm-1 resolution. Thirty-two scans were collected per spectrum. A total of five hundred sixty-two samples were collected. 80% of the samples were used to build calibration models using the NIR spectra and reference values and 20% were used as a validation test set.
| Acidity (AV) | R² = 0.96 |
| ρ-Anisidine (pAV) | R² = 0.95 |
| Total Polar Materials (TMP) | R² = 0.99 |
| Peroxide Value (PV) | R² = 0.93 |
| Oxidative Stability Index (OSI) | R² = 0.91 |
Correlation coefficients were high for all parameters and validation set predictions proved the validity of the calibration models. Some samples did show chemical anomalies in the reference testing, and before implementing these calibrations in a real-time setting, more such samples will be needed in the models. The study showed the potential to replace traditional methods for monitoring thermo-oxidative degradation in frying oils using NIR spectroscopy.
https://journals.sagepub.com/doi/pdf/10.1177/0003702818764125
Online Monitoring of P(3HB) Produced from Used Cooking Oil with Near-Infrared Spectroscopy – Cruz, Sarraguca, Freitas, et al., Journal of Biotechnology 194(2015) 1-9
Polyhydroxyalkanoates (PHA) are polyesters produced in nature by numerous microorganisms, including through bacterial fermentation of sugar or lipids. The simplest and most commonly occurring form is the fermentative production of poly-3-hydroxybutyrate (P3HB). There is high interest in the creation of PHA-based materials on an industrial scale because they are biodegradable while having properties of plastics as well as a way to create plastics from non-fossil fuel sources. Oil containing feedstocks are used as alternative substrates to glucose and sucrose for PHA production because high conversion yields can be obtained. Used cooking oil is one option for the substrate. A batch reactor was operated producing P(3HB) using used cooking oil as the sole carbon source. NIR spectroscopy was used for online monitoring of the fermentation. Spectra were collected using a transflectance probe from 10000 cm-1 to 4000 cm-1. Resolution was 8 cm-1 and sixteen scans were collected and averaged per spectrum. Samples were pulled from the reactor and reference tests were performed for Biomass, Used Cooking Oil (UCO), and Polyhydroxyalkanoates (PHA).
| Biomass | R² = 0.86 |
| Used Cooking Oil (UCO) | R² = 0.96 |
| Polyhydroxyalkanoates (PHA) | R² = 0.78 |
The results prove the feasibility of using NIR spectroscopy and calibration models as an on-line monitoring tool for Biomass, UCO, and PHA. This study was the first to successfully use NIR as a method for monitoring these specific parameters in a bioreactor. Further work will be needed for full implementation of the method but it was validated as a way to estimate these three important parameters with no significant analytical, operational costs.
https://www.sciencedirect.com/science/article/pii/S0168165614010207
The post Cooking Oil Analysis appeared first on NIR-For-Food.
]]>